Hemolysis associated with pneumatic tube system transport for blood samples
Abstract
Objective: The frequency of hemolysis of blood samples may be increased by transport in a pneumatic tube system. The purpose of this study was to evaluate the effect of pneumatic tube system transport on hemolysis of blood samples.
Methods: Blood samples were transported from the emergency department to the hospital laboratory manually by hospital staff (49 patients) or with a pneumatic tube system (53 patients). The hemolysis index and serum chemistry studies were performed on the blood samples and compared between the different methods of transport.
Results: The blood samples that were transported by the pneumatic tube system had a greater frequency of hemolysis and greater mean serum potassium and median creatinine, aspartate aminotransferase, and lactate dehydrogenase levels than samples transported manually.
Conclusion: Blood samples transported from the emergency department to the hospital laboratory by a pneumatic tube system may have a greater frequency of hemolysis than samples transported manually. This may necessitate repeat phlebotomy and cause a delay in completing the laboratory analysis.
doi: http://dx.doi.org/10.12669/pjms.301.4228
How to cite this:Kara H, Bayir A, Ak A, Degirmenci S, Akinci M, Agacayak A, et al. Hemolysis associated with pneumatic tube system transport for blood samples. Pak J Med Sci 2014;30(1):50-53. Â doi: http://dx.doi.org/10.12669/pjms.301.4228
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.